Description
About:
- Autozyme Albumin is a reagent set for determination of albumin in serum or plasma based on BCG method
- Autozyme Albumin is a single reagent system, ready to use.
- Autozyme Albumin is linear up to 6 gm%
- Albumin can be determined in 1 minute at room temperature
- The shelf-life of Autozyme Albumin is 18 months
Pack Size : 2 x 50 ml
Principle:
Serum albumin in the presence of Bromocresol-green under acidic condition forms a green coloured complex, The absorbance of this complex is proportional to the albumin concentration in serum or plasma.
Procedure:
Reaction type ………………. End-Point
Reaction time …………. ………….1 min. at R.T. (25-30°C)
Wavelength …………. …………….. 600 nm.(600 – 650 nm.)
Zero setting with …………………. Reagent Blank
Blank absorbance limit …… ………….. < 0.100 Abs.
Sample volume …………….0.01 ml (10 ul)
Reagent volume …………. …………. 1.0 ml
Standard concentration ……………. 5 gm%
Linearity. ………….. 6 gm%
Quality Control:
To ensure adequate quality control, it is recommended that each batch should include a normal and an abnormal commercial reference control serum,. it should be realized that the uSe of quality control material checks both instrument and reagent functions together. Factors which might affect the performance of this test include proper instrument function, cleanliness
of glassware and accuracy of pipetting.
The reagents are for in vitro diagnostic use.
BCG reagent and standard should be stored at temperature indicated on the bottle label.
Reviews
There are no reviews yet.