Description
Autochem Ingeniious is a semi automated clinical chemistry analyser & has been designed to meet the demands of the research as well as laboratory staff. Compact ergonomic design which is ideal for use in small-to-medium-sized laboratory.
- In-built Peltier regulated incubation compartment
- Integrated Coagulation Testing Mode
- Single & profile wise testing facility
- Flexible calibration modes
- Direct access key to run tests
- Enhanced Patient Data Storage & Retrieval
- Daily & monthlty QC program
- Wash & Aspiration function can be controlled using External Keyboard (PS/2 keyboard)
Technical Specifications
OPTICAL SYSTEM / PHOTOMETER
Optical Design: Single beam with rotating filter wheel
Measurement Mode: Monochromatic & Bichromatic
Light Source: Quartz Halogen Lamp, 12 V, 20 W
Photometric Range: -0.005 – 2.5 Absorbance
Standard wavelengths: 340, 405, 510, 546, 578, 630 nm + Two optional
Filter Type: 4-Cavity interference, long life ion beam assisted deposition
Wavelength accuracy: +/- 5nm
Sample Holder: 18 µL flow cell, 10 mm square/semi-micro cuvette, Coagulation adaptor and cuvette
DATA PROCESSING
Calculation Modes
-
Absorbance
-
End Point (with & without sample blank)
-
Kinetic (with and without Reagent Blanking)
-
Fixed Time (with and without Reagent Blanking)
-
1-Point with Sample Blank
-
Coagulation
-
Profile (up to 8 tests/profile)
Calibration
-
Factor mode
-
Single-standard mode
-
Multi-standard mode (up to 6 standard) (spline)
-
Multi-standard mode (up to 6 standard) (point to point)
Data storage capacity
-
144 test parameters
-
24 profiles
-
2000 (latest) samples, 200 calibrators/standards, 576 daily controls
-
10 (latest) graphs
Result retrieval: Patient ID wise, Test wise, Date wise
TEMPERATURE CONTROL
Cuvette temperature: Peltier based
Operating temperature: 30 & 37 °C (+/- 0.2 °C)
Incubation block: Inbuilt for 37°C ± 0.2°C
-
16 places for 12 mm diameter test tubes
FLUIDIC SYSTEM
Flow cell volume: 18 µL
Aspiration volume: 200 µL – 2000 µL
Aspiration: Stepper driven peristaltic pump
Data Management
-
Display type: 240 x 128 resolution graphical LCD with backlight
-
Data input: In-built 28 keys membrane keypad
-
Facility to connect to external standard computer keyboard
-
Real time curve plotting of absorbance Vs. Time
-
Internal thermal printer (speed up to 0.5 lines/second)
-
Facility to connect to external 80 column dot matrix printer
-
Serial port for computer connectivity
POWER / DIMENSIONS
Power requirements: Input Voltage : 90-260 VAC, 50/60 Hz
Power Consumption: 100 Watts Max
Operating Environment: 20°C to 40°C, RH: 85% max
Relative Humidity: 10% to 85%, Non-condensing
Dimensions: 35 cm (W) x 34 cm (D) x 26 cm (H)
Weight: < 10 kg
In-built Peltier regulated incubation compartment
Cost Efficient
-
Saves cost of buying additional Incubator (approx. Rs.5000)
-
Saves space of approx. 1 Sq. Feet for placing the additional incubator
Flexible
-
On/Off facility for using incubator when required
Convenient
-
Integrated timer with buzzer for ensuring right incubation time
Integrated Coagulation Testing Mode
Cost Efficient
-
Saves cost of buying additional Coagulation Analyser (approx. Rs.75,000)
-
Saves space of approx. 1 Sq. Feet for placing the additional incubator
-
No manual calculation of INR
Convenient
-
No maintenance of additional coagulation instrument & accessories
(Image: Special Coagulation Cuvette, Coagulation Cuvette Holder)
Single & Profile Wise Testing Facility
Program to create 24 different profiles / panels & upto 8 different tests in each profile
-
Convenience of running multiple parameters on same sample without switching program
-
Enables faster sample turn around
Flexible Calibration Modes
Single & Multi-Point Calibration Possible
-
Flexibility to run routine as well as specialized parameters (Eg. Immunoturbidimetry) on same instrument
Auto-normalization of Calibration curve making recalibration simple & cost effective
-
Recalibration can be done using any 1 Standard and the instrument will automatically normalize the calibration curve
-
Save cost of expensive reagents and standards
Easy Operation
-
Wash & Aspiration function can be controlled using External Keyboard (optional PS/2 keyboard)
-
Direct access key to run tests
-
Saves time by directly running 9 program tests with single key from main menu, without accessing detailed test menu
-
Reduces chance of error by eliminating the need to memorize test menu number
-
Enhanced Data Management Capability
Enhanced Patient Data Storage & Retrieval
-
Stores data of latest 2000 samples, 200 calibrators/standards, 576 daily controls and 10 reaction graphs
-
Flexibility to retrieve Sample & Calibration results by – Patient ID, Date & Test Name
-
Facility to input patient information & demographic details
Daily & monthly QC program
-
Manage 2 levels of controls with Levey-Jennings Chart
Reviews