Each batch of AutoPure AST (GOT) is assayed with at least six quality control sera — Precinorm**, Precipath**, Biorad l***, Biorad II***, Sero Low**** and Sero High****prior to release.
To ensure adequate quality control, it is recommended that the laboratory should use a normal and an abnormal commercial reference control serum.
It should be realized that the use of quality control material checks both reagent and instrument functions together. The value of these controls should fall within the specified limits. If control values fall outside specified limits, each of the below criteria should be cross-checked and corrected :
• Proper instrument function – wavelength setting, light source and
temperature control
• Cleanliness of probes & cuvettes
• Bacterial contamination of wash water used by the instrument
• Expiry date of the reagent kit
Reviews
There are no reviews yet.