Description
About:
- Autozyme Bilirubin is a reagent set for determination of Total and Direct Bilirubin based on Jendrassik & Grof method, using Diazotized Sulphanilic Acid with DMSO as an activator for Total Bilirubin
- Autozyme Bilirubin is a two reagent system using one procedure
- Autozyme Bilirubin is linear up to 20mg%
- Autozyme Bilirubin has one step reconstitution
- Autozyme Bilirubin can be determined in just 5 minutes
- Autozyme Bilirubin method is relatively free of interference from haemoglobin and other commonly occuring substances in serum or plasma
Pack Size : 2 x 100 ml
Principle:
Bilirubin reacts with diazotized sulfanilic acid to produce azobilirubin (violet color). DMSO catalyzes the formation of azobilirubin from free bilirubin, The violet color is proportional to bilirubin concentration measured at 546 nm.(530 – 550 nm.)
Procedure:
Reaction type…………………. End-Point
Reaction time………………….5 min at R.T (25-30°C)
Wavelength………………….546 nm. (530-550 nm)
Zero setting with………………….Serum Blank
Sample volume………………….0.05 ml (50 µl)
Reagent volume………………….1.02 ml
Factor………………….20.2
Linearity………………….20 mg%
Quality Control:
It’s Recommended that each batch should include a normal and an abnormal commercial reference control serum or a known patient serum, Use of quality control serum, checks both instrument and reagent functions together. Factors which might affect the performance of this test include proper instrument function, temperature control, cleanliness of glassware. Accuracy of pipetting and serum exposure to light.
The reagent kit should be stored at Room Temperature (25-30°C) and is stable till the expiry date indicated on the label.
The working solution is stable for 2 days at 2- 80 and for 5 hours at Room Temperature (25-30°C).
Reviews