Description
About:
- Autozyme GOT is a reagent set for determination of GOT based on UV-Kinetic method recommended by IFCC
GOT activity can be determined in 2.5 minutes - Autozyme GOT is linear up to 400 IU/I and has high stability
- Autozyme GOT is a single reagent system, using one step procedure
Principle:
The conversion of NADH to NAD is proportional to the concentration of GOT n serum and is measured at 340 NM as rate of decrease in
absorbance.
Quality Control:
To ensure adequate quality control, it is recommended that each batch should include a normal and abnormal commercial reference control serum.
It should be realized that the use of quality control material checks both instrument and reagent functions together. factors which might affect the
performance of this test include proper instrument function, temperature Control. Cleanliness of glassware, and accuraCy of pipetting.
The reagent kit should be stored at 2-8°C and is stable till the expiry date indicated on the label.
The GOT working solution is stable for 60 days at 2-8°C.
Do not freeze the working solution.







Reviews
There are no reviews yet.